1.3The method's purpose sets the target
What is happening
The test type decides what resolution has to be achieved, over what dynamic range, and against what interferences — before any chemistry is chosen.
Worth memorising
Assay: accuracy near 100 %. Impurities: sensitivity and specificity near 0.05 %. You cannot optimise for both at once without knowing you are doing it.
At the bench
Write the test type at the top of the development plan. Every acceptance number in the plan should be traceable to it.
What the test type changes
| Test | The hard part | What development must prove | Typical range |
|---|---|---|---|
| Assay (content) | Accuracy and precision at one level | The API peak is free of anything that co-elutes and absorbs | 80–120 % of nominal |
| Related substances | Seeing and resolving small peaks beside a large one | Every known and stress-generated degradant is resolved and detectable at the reporting threshold | Reporting threshold to ~120 % of the limit |
| Dissolution | Many samples, fast, in a surfactant-bearing medium | The method survives the medium and runs quickly enough to be practical | Typically ~10–120 % of label |
| Identity | Discrimination, not quantity | Retention (and spectrum, if available) distinguishes the API from plausible alternatives | Not applicable |
The resolution target is a decision, not a default
A baseline-resolved pair of equal peaks is usually described as Rs 1.5. That figure assumes Gaussian peaks of similar size. When one peak is two hundred times the other — an impurity beside an API — the tail of the large peak sits under the small one long before the stated resolution says they are separated. For critical pairs of very unequal size, developers commonly target higher resolution than 1.5, and it is worth deciding that number deliberately rather than inheriting it.