The method's purpose sets the target
Part 1 · Strategy & Foundations · 8 min
Part 1 · Strategy & Foundations

1.3The method's purpose sets the target

An assay and a related-substances method are different instruments wearing the same conditions.

What is happening

The test type decides what resolution has to be achieved, over what dynamic range, and against what interferences — before any chemistry is chosen.

Worth memorising

Assay: accuracy near 100 %. Impurities: sensitivity and specificity near 0.05 %. You cannot optimise for both at once without knowing you are doing it.

At the bench

Write the test type at the top of the development plan. Every acceptance number in the plan should be traceable to it.

What the test type changes

TestThe hard partWhat development must proveTypical range
Assay (content)Accuracy and precision at one levelThe API peak is free of anything that co-elutes and absorbs80–120 % of nominal
Related substancesSeeing and resolving small peaks beside a large oneEvery known and stress-generated degradant is resolved and detectable at the reporting thresholdReporting threshold to ~120 % of the limit
DissolutionMany samples, fast, in a surfactant-bearing mediumThe method survives the medium and runs quickly enough to be practicalTypically ~10–120 % of label
IdentityDiscrimination, not quantityRetention (and spectrum, if available) distinguishes the API from plausible alternativesNot applicable

The resolution target is a decision, not a default

A baseline-resolved pair of equal peaks is usually described as Rs 1.5. That figure assumes Gaussian peaks of similar size. When one peak is two hundred times the other — an impurity beside an API — the tail of the large peak sits under the small one long before the stated resolution says they are separated. For critical pairs of very unequal size, developers commonly target higher resolution than 1.5, and it is worth deciding that number deliberately rather than inheriting it.